- Description
Features:
1. Universal syringe set: 5ml,10ml,20ml,30ml,50ml,60ml.
2. Automatically detection of syringe types and remained solution.
3. Broad range in delivery rate settings.
3. 6 Hours continuous work after fully charged.
4. Three levels of alarm volume: high, medium and low.
5. Three levels of occlusion pressure: high, medium and low.
6. KVO function:After infusion is complete, the pump automatically switches to the smallest flow to prevent creating the thrombus.
OEM Service Available
CE/ISO13485 Marked
Technical Specifications
1. Universal syringe set: 5ml,10ml,20ml,30ml,50ml,60ml.
2. Automatically detection of syringe types and remained solution.
3. Broad range in delivery rate settings.
3. 6 Hours continuous work after fully charged.
4. Three levels of alarm volume: high, medium and low.
5. Three levels of occlusion pressure: high, medium and low.
6. KVO function:After infusion is complete, the pump automatically switches to the smallest flow to prevent creating the thrombus.
OEM Service Available
CE/ISO13485 Marked
Technical Specifications
PARAMETER | SPECIFICATION |
Applicable syringes | 10ml,20ml,30ml,50ml,60ml of any standard |
Accuracy | +/- 2% with correct calibration |
Rate mode | 50ml(60ml):0.1ml/h-1800.0ml/h (0.1ml/h step) 20ml(30ml):0.1ml/h-800.0ml/h (0.1ml/h step) 10ml:0.1ml/h-400.0ml/h (0.1ml/h step) |
Time Mode | SOL. VOL:0.1-999.9ml (0.1ml step) TIME:1-1999min (1min step) |
Body Weight Mode: | Volume:0.1-999.9ml(0.1ml step) DRUG MASS:0.1-999.9mg 0.1mg step) DOSE:0.01-99.99ug/kg/min(0.01ug/kg/min step) WEIGHT:0.1-300kg(0.1kg the least step) |
Total Volume Range | 0.1-1999.9ml (0.1 ml step) |
Purge Rate | 50ml(60ml):1800.0ml/h, 20ml(30ml):800.0ml/h, 10ml:400.0ml/h |
KVO Rate | 1ml/h |
Occlusion pressure | Three level setting LOW, MEDIUM or HIGH |
Alarms | Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low battery, device failures. Three levels of alarm volume LOW, MEDIUM or HIGH |
Bolus Rate | 10ml:10-300ml/h,20ml(30ml):10-600ml/h, 50ml(60ml):10-1200ml/h |
Bolus Volume | Maximum of 25ml |
Interface | RS-232, Reserved port |
Power source | AC power supply: AC 100 to 240V( +/-10%),50/60Hz; Built-in Lithium Battery: Rechargeable, 11.1V, 2000mAh. 6 hours of continuous operation. |
Accessories: | AC power line, Operational manual |
Protection Class | Class I, Type CF,IPX4 |
Dimension / Net weight | 300mm*240mm*250mm , 3.6kg |
Shipping Information:
G.W: 4.8kg
Packing Size: 0.34m,0.32m,0.26m
Unit: Piece
Special: No
Name:
Email:
Whatsapp/Tel:
Message: