- Description
- USER MANUAL
- Detail pictures
Specifications
Type:
Trolley Type
Ventilation Mode:
IPPV, SIPPV, IMV, SIMV, VCV, PEEP,STANDBY, MANUAL
Main technical specifications:
Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from O2 & NO2: 0.32~0.6 MPa.
5-tubes Flowmeters, O2 & N2O & Air ; 0.1L/Min ~ 10L / Min;
Oxygen Flush: 25~75 L / Min
Respiration Mode: A/C,IPPV,SIPPV,IMV,SIMV,VCV,PEEP,SIGH,MANUAL (Optional: PCV )
Tidal Volume: Adjustable range: 50 ~ 1500 ml, Display range:0~2000ml
I/E Ratio: 8:1 ~ 1:10
Two vaporizers: Enflurane/ Isoflurane/Sevoflurane/Halothane
Vaporizer Setting Range: 0.5%~5%
Respiration Frequency : 1~100 bpm;
Lung compliance: ≤30 mL / kPa
IMV frequency: 1~20 bpm;
PEEP: 0 ~ 2.0 kPa
Ptr: -1.0kPa ~ 2.0 kPa
CO2 absorber capacity : 1 kg
Inspiratory Plateau: 0~1 S ;
O2 Concentration: 21%~100%
Optional: Patient monitor
Alarm parameters:
upper/lower limit of airway pressure,tital volume exceed limitation,no tidal volume output,apnea,oxygen concentration, AC power malfunction,Battery low power.
Standard Configuration:
Machine with two Vaporizer,
(please choose from Enflurane/ Isoflurane/Sevoflurane/Halothane)
Optional:
Patient Monitor
Shipping Information:
G.W: 114kg
Packing Size: 0.94m,1.48m,1.23m
Unit: Piece
Special: No