- Description
Features:
1. Compatible with syringes of any brand, 10ml~60ml.
2. Automatically calibration for syringes of any brand.
3. Three working modes: Rate control mode, Time control mode and Dose-Weight control mode.
4. Automatically detection of syringe types and settings.
5. Wide range in delivery rate settings.
6. 6 Hours continuous worked.
7. Three levels of alarm volume: high, medium and low.
8. Three levels of occlusion pressure: high, medium and low.
9. KVO function:After infusion is complete, the pump automatically switches to the smallest flow to prevent creating the thrombus.
OEM Service Available
CE Marked
Technical Specifications
1. Compatible with syringes of any brand, 10ml~60ml.
2. Automatically calibration for syringes of any brand.
3. Three working modes: Rate control mode, Time control mode and Dose-Weight control mode.
4. Automatically detection of syringe types and settings.
5. Wide range in delivery rate settings.
6. 6 Hours continuous worked.
7. Three levels of alarm volume: high, medium and low.
8. Three levels of occlusion pressure: high, medium and low.
9. KVO function:After infusion is complete, the pump automatically switches to the smallest flow to prevent creating the thrombus.
OEM Service Available
CE Marked
Technical Specifications
PARAMETER | SPECIFICATION |
Applicable syringes | 10ml,20ml,30ml,50ml,60ml of any standard |
Accuracy | +/- 2% with correct calibration |
Rate mode | 50ml(60ml):0.1ml/h-1800.0ml/h (0.1ml/h step) 20ml(30ml):0.1ml/h-800.0ml/h (0.1ml/h step) 5ml,10ml:0.1ml/h-400.0ml/h (0.1ml/h step) |
Time Mode | SOL. VOL:0.1-999.9ml (0.1ml step) TIME:1-1999min (1min step) |
Body Weight Mode: | Volume:0.1-999.9ml(0.1ml step) DRUG MASS:0.1-999.9mg(0.1mg step) DOSE:0.01-99.99ug/kg/min(0.01ug/kg/min step) WEIGHT:0.1-300kg(0.1kg the least step) |
Total Volume Range | 0.1-1999.9ml (0.1 ml step) |
Purge Rate | 50ml(60ml):1800.0ml/h, 20ml(30ml):800.0ml/h, 10ml:400.0ml/h, 5ml: 200ml/h |
KVO Rate | 1ml/h |
Occlusion pressure | Three level setting LOW, MEDIUM or HIGH |
Alarms | Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low battery, device failures. Three levels of alarm volume LOW, MEDIUM or HIGH |
Bolus Rate | 10ml:10-300ml/h,20ml(30ml):10-600ml/h, 50ml(60ml):10-1200ml/h |
Bolus Volume | Maximum of 25ml |
Interface | RS-232 |
Power source | AC power supply: AC 100 to 240V( +/-10%),50/60Hz; Built-in Lithium Battery: Rechargeable, 11.1V, 2000mAh. 6 hours of continuous operation. |
Accessories: | AC power line, Operational manual |
Protection Class | Class I, Type CF,IPX4 |
Dimension / Net weight | 280mm*160mm*130mm , 2.5kg |
Shipping Information:
G.W: 3kg
Packing Size: 0.26m,0.2m,0.18m
Unit: Piece
Special: No